TNP #39 - Faxitron MX-20 Digital Micro-Focus X-Ray Machine Teardown, Repair & Experiments
The Signal Path The Signal Path
121K subscribers
25,204 views
0

 Published On Sep 17, 2023

In this episode Shahriar repairs his Faxitron MX-20 digital x-ray system. Although the machine has been producing beautiful images for years, recently it has been outputting out-of-focus images intermittently.
The principle of x-ray focusing and beam-spot generation is presented along with potential electronic methods of achieving focus. The teardown of the unit reveals the main x-ray tube controller PCB which houses the focus control high-voltage modules. Since the problem is intermittent, a reed switch is also suspected. The high-voltage module is measured and appears fully functional. Replacing the reed switch corrects the out-of-focus images. Several x-ray images showcases the sharpness of the unit after the repair.